Flare
This metric is the best area of Imatest support of IEEE 2020-2024.
Method A
Similar to ISO 18844 but potentially within a hemisphere to support wider field
Chart: 2020-2024 requires more black patches than the 18844 method C chart. These patches can be arranged within on a dome to help with wide FOV testing.
We support the P2020 flare method A from the calculation side, although dot proximity could lead to some issues if the dots are too close together.
Method B
This is what Imatest’s stray light (flare) solution aims to solve, as we think the second method is a more comprehensive test. Our stray light solution contains a collimated LED source, a Motorized Gimbal, as well as flexible software analysis. Our solution can be configured to support the second method of the IEEE 2020-2024 standard.
Geometry
We support geometric calibration validation as a test lab service. We have a partner Furonteer that offers production level geometric calibration machines.
Geometric calibration is not (and will not be part) of the standard.
Geometric calibration validation is a part of the standard. For some calibration models, we can do this as a service.
Noise
We support some of the metrics within the P2020 noise standard, and plan to update Imatest software to support the rest of the metrics in the medium term.
Imatest supports basic means and standard deviations, similar to the standard.
We do not support "Raw Video" which may be required for some of the temporal analysis.
Dynamic Range
To support IEEE 2020-2024 calculation, we will ensure we have the right chart patterns, which may be new patterns, as our software already has the necessary calculations available.
Many members of IEEE P2020 used the contrast resolution chart.
Additional analysis code to support the standard will be added in the medium term.
SFR
The charts for SFR are a slanted square surrounded by registration marks and tonal patches, so local linearity and magnification can be obtained. Imatest does not yet produce this target.
IEEE 2020 SFR calculation uses “ISO 12233:2024" slanted edge calculation.
Our gamma calculation may be superior to the local linearity calculations, especially in the cases of extreme non-uniformity.
Flicker
We currently sell hardware to test flicker per P2020 in the form of a flickering light box, but our software does not yet support the calculations. See http://imatest.com/imaging/flicker/
In lieu of software support, we could perform flicker analysis for you on a consulting basis within the context of our lab services, contact labservices@imatest.com.
Contrast Performance Indicator (CPI)
Our contrast resolution and HDR charts can be utilized to capture images that can support this standard, and we have some capability to support multi-image analysis that we can expand upon if needed.
Contrast Transfer Accuracy (CTA) was formerly known as CDP “Contrast Detection Probability”.
Contrast Signal to Noise Ratio is based upon this paper by Robin Jenkins presented at Electronic Imaging in 2021, the permission to use this NVIDIA patent has supposedly been granted.
We released a CDP calculation in Imatest 5.1 based on an early draft, which is not IEEE 2020 CTA. We will have to update things to get compliant.
Related links
Purchase the standard: https://www.techstreet.com/ieee/standards/ieee-p2020?gateway_code=ieee&vendor_id=6765&product_id=2505612
Imatest’s standard page: https://www.imatest.com/imaging/ieee-2020/